Volcano s5 Imaging system, Ultrasound pulsed echo imaging system. Product consists of the CPU, control console, monitor, printer, and patient interface module (PIM)

FDA Recall Z-1190-2007 · Posted 2007-08-23

Recalling FirmVolcano Corporation
Recall StatusTerminated
Product CodeIYO
Date Initiated2007-04-02
Date Terminated2008-06-03
Product Quantity351 units worldwide

Reason for Recall

During emissions testing, it was discovered that the product exceeds emissions limits for electromagnetic compatibility.

Firm Action

The firm notified its customers by letter beginning April 2, 2007, and started replacing the monitor with appropriate replacement parts.

Root Cause

Other

Distribution Pattern

Product was distributed to health care facilities nationwide and to distributors in Europe (via subsidiary), Japan (via subsidiary), Latin America and Asia (third party distributors).