Steris C1160 Universal Flexible Processing Tray, for use with STERIS SYSTEM 1 Processor, used for sterilization of flexible endoscopes.
FDA Recall Z-1190-2008 · Posted 2008-08-05
Advertisement
Recalling FirmSteris Corp
Recall StatusTerminated
Product CodeMED
Date Initiated2008-02-19
Date Terminated2012-02-24
Product Quantity28,766 units
Reason for Recall
Failure to effectively sterilize: Under certain conditions, the design/operation of the C1160 Tray may compromise the ability of the SYSTEM 1 processor to correctly assess the state of the high pressure pump during the Sterile Cycle. The L6 sensor fails to alarm when the high pressure pump is not working properly.
Firm Action
Steris sent an Urgent Safety Alert/Recall letter dated 2/8/2008, to all customers. The letter stated that any device sterilized in the C1160 Tray should not be used clinically prior to a completion of a successful diagnostic cycle. Until further notice, you must run a diagnostic cycle after each sterile cycle involving a C1160 tray or alternately you must discontinue use of these trays. Steris service technicians will schedule visits to correct the problem.
Root Cause
Device Design
Distribution Pattern
Worldwide Distribution, including USA, Puerto Rico, Canada, and Japan.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement