BALLARD* ACCESS Closed Suction System for Neonates/Pediatrics, Elbow Manifold.
FDA Recall Z-1190-2023 · Posted 2023-03-13
Advertisement
Recalling FirmAvanos Medical, Inc.
Recall StatusTerminated
Product CodeBSY
Date Initiated2023-02-03
Date Terminated2024-09-05
Product Quantity220 units
Reason for Recall
Potential for product manifold to crack during use on patients, possibly causing leaks in the respiratory circuit and can lead to the patient receiving inadequate ventilation.
Firm Action
Avanos hand-delivered the recall notification to the single consignee on 02/03/2023. The letter instructed the customer to discontinue use of and quarantine their stock of all lots of affected units and return them. Additionally, the customer was instructed to complete and return the acknowledgement form.
Root Cause
Component design/selection
Distribution Pattern
US Nationwide distribution in the state of Oklahoma.