LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a modular stem coupled with a proper neck by means of a Morse taper stabilized during the implantation phase by a safety screw. This system is particularly indicated for revision surgery on both uncemented and cemented femoral implants, when there is significant bone loss and an abnormal meta-epiphyseal anatomy of the femur.
FDA Recall Z-1191-2017 · Posted 2017-02-11
Advertisement
Recalling FirmEncore Medical, Lp
Recall StatusTerminated
Product CodeKWA
Date Initiated2017-01-10
Date Terminated2017-03-31
Product Quantity13 devices
Reason for Recall
Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma sterilization.
Firm Action
DJO Global sent an Urgent Field Safety Notice letter dated January 10, 2017 to customers. The letter identified the affected product, problem and actions to be taken. Customers were asked to complete the Acknowledgement and receipt Form. Customers were instructed to contact Customer Service at 1-800-456-8696 to explore options for a replacement order.
Root Cause
Nonconforming Material/Component
Distribution Pattern
US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.