NxStage Cartridge Express with Speedswap, Catalog No. CAR-535. An accessory device for facilitating the removal and replacement of a flow-compromised filter during treatment of acute and chronic renal failure or fluid overload.

FDA Recall Z-1191-2024 · Posted 2024-02-27

Recall StatusOpen, Classified
Product CodeKDI
Date Initiated2024-01-29
Product Quantity693

Reason for Recall

Product is recalled due to reports of blood leaking or spraying from the arterial line on the cartridge.If this failure occurs, a patient may experience uncontrolled blood loss or a healthcare professional could be exposed to infectious patient blood.

Firm Action

The firm began notifying affected customers on January 29, 2024. Customers were informed that the issue may occur when the needleless access port located next to the filter has been accessed and there is a buildup of pressure in the circuit. Customers were instructed to inspect their inventory for affected product and quarantine recalled product. Customers should not use affected product and should ensure all users of the CAR-535 are aware of the notice. Customers should contact NxStage Customer Service at 1-866-NXSTAGE (1-866-697-8243) to arrange for the return of all affected products. This number may also be contacted if you have any questions or comments.

Root Cause

Component design/selection

Distribution Pattern

US domestic distribution to Dialysis Centers located in AZ, CA, CO, GA, IL, IN, MA, MD, MI, MN, MO, MS, NC, NJ, NY, TN, TX, VA, WI. No international distribution.

Related FDA 510(k) Records