Open, Classified

GRAPHENANO DENTAL G-CAM, Product Codes: a) 95 Multichroma 14LB2, REF GC00089; b) 95 Multichroma 24LA2, REF GC00112; c) 95 Multichroma 20LA1, REF GC00130; d) 95 Multichroma 24LC2, REF GC00090; e) 95 Multichroma 24LA1, REF GC00113; f) 95 Multichroma 16LB2, REF GC00131; g) 95 Multichroma 14LA2, REF GC00157; h) 95 Multichroma 20LB2, REF GC00098; i) 95 Multichroma 20LBL2, REF GC00114; j) 95 Multichroma 16LA3,5, REF GC00132; k) 95 Multichroma 14LA1, REF GC00158; l) 95 Multichroma 24LB1, REF GC00099; m) 95 Multichroma 20LA3,5, REF GC00115; n) 95 Multichroma 16LA3, REF GC00133; o) 95 Multichroma 16LC2, REF GC00091; p) 95 Multichroma 14LBL2, REF GC00134; q) 95 Multichroma 24LBL2, REF GC00092; r) 95 Multichroma 24LA3,5, REF GC00101; s) 95 Multichroma 20LA3, REF GC00117; t) 95 Multichroma 14LB1, REF GC00135; u) 95 Multichroma 24LB2, REF GC00094; v) 95 Multichroma 20LA2, REF GC00119; w) 95 Multichroma14LA3,5, REF GC00137; x) 95 Multichroma 14LC2, REF GC00095; y) 95 Multichroma 24LA3, REF GC00104; z) 95 Multichroma14LA3, REF GC00138 ; used in the manufacture of dentures.

FDA Recall Z-1192-2024 · Posted 2024-02-28

Recalling FirmGRAPHENANO DENTAL SL
Recall StatusOpen, Classified
Product CodeEBI
Date Initiated2023-10-24
Product Quantity354 units

Reason for Recall

Machining error on cams renders the device unusable.

Firm Action

Graphenano Dental notified its sole consignee on 10/24/2023 via email. The notice explained the problem with the device and requested that the consignee quarantine all affected devices pending further instructions. A Graphenano employee examined all quarantined product at the consignee for defective devices and were subsequently destroyed.

Root Cause

Process control

Distribution Pattern

US Nationwide distribution in the state of Utah.

Related FDA 510(k) Records