Stryker VertePort, 10 G Cement Cannula, REF 306-410, Qty 18; For use with the Stryker PCD Precision System and ACM Delivery System, Rx only, sterile, Stryker Instruments, Kalamazoo, MI.

FDA Recall Z-1193-2008 · Posted 2008-03-11

Recall StatusTerminated
Product CodeJDZ
Date Initiated2008-02-13
Date Terminated2008-12-30
Product Quantity953 boxes Recalls Z-1192-1195-2008

Reason for Recall

Mis-labeling; The instructions for use include a use for which the device is not intended. (Device not approved to capture a bone biopsy)

Firm Action

Consignees were notified via Medical Device Recall Notification letter dated 2/13/08 to read the enclosed instructions for use, remove all old instructions for use at the consignee location and replace them with the new instructions for use, and to discard the old instructions for use.

Root Cause

Labeling Change Control

Distribution Pattern

Nationwide.