Terminated

Stryker Restoris RIO Reamer Handle, Offset, Mako Rx only Product Usage: The handle is used to hold surgical components during acetabular reaming in MAKO Total Hip procedures.

FDA Recall Z-1193-2017 · Posted None

Recall StatusTerminated
Product CodeOLO
Date Initiated2016-09-29
Date Terminated2019-08-06
Product Quantity595 units

Reason for Recall

Failure of the Offset Cup Reamer Handle not engaging its mating components.

Firm Action

Customers were sent a recall notificaiton letter, dated 10/17/2016, addressing the reason for recall and the appropriate actions that should be taken by customers. Consignees are instructed to return any recalled product on hand to Stryker C/O Stericycle, 2670 Executive Dr. Suite A, Indianapolis, IN 46241, Attn RA 2016-083, RA 2016-119 - Event 8811. Consignees are also instructed to inform their users of this recall and forward the notice to affected individuals. A Business Reply Form was provided and is to be completed within 5 business days of receipt by consignees. Completed business reply forms are to be faxed to 855-620-5693 or emailed to [email protected].

Root Cause

Component change control

Distribution Pattern

Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NJ, NM, NY, OH, OK, OR, PA, RI, TN, TX, UT, VA, WA, WI, WV, & WY and the countries of Italy, Germany, Australia, China, Europe, Hong Kong, South Africa, Germany, Thailand & Singapore.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.