ARROW ENDURANCE" Extended Dwell Peripheral Catheter System, Catalog Numbers: EDC-00820 & EDC-00820-B

FDA Recall Z-1194-2018 · Posted 2018-03-14

Recall StatusTerminated
Product CodeFOZ
Date Initiated2017-12-15
Date Terminated2019-07-18
Product Quantity305

Reason for Recall

Certain lots of peripheral catheter systems may be missing the safety latch, intended to minimize the risk of sharps injuries.

Firm Action

Notification letters were sent to consignees with the following instructions: 1. If you have affected stock, quarantine so that the affected products can be returned to Arrow. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to Arrow. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to Arrow.

Root Cause

Process control

Distribution Pattern

Distributed in the following U.S. states: CO, FL, IL, NC, PA, SC, and TX.