Nucleic Acid test intended for screening for MRSA/SA. Each kit has 20 tubes. in vitro diagnostic.
FDA Recall Z-1195-2018 · Posted None
Advertisement
Recalling FirmRoche Molecular Systems, Inc.
Recall StatusTerminated
Product CodeNQX
Date Initiated2017-10-27
Date Terminated2020-04-27
Product Quantity136 kits
Reason for Recall
The cobas MRSA/SA tests may have decreased performance compared to the analytical sensitivity for MSRA detection. A screening failure can lead to potential infection in the colonized patient being tested as well as the spread of MSRA to others. The United States is not impacted, as the cobas MRSA/SA nucleic acid test for use on the cobas liat system is not available in the US.
Firm Action
The cobas MRSA/SA tests has the potential for decreased performance compared to the analytical sensitivity for MSRA detection. A screening failure can lead to potential infection in the colonized patient being tested as well as the spread of MSRA to others.
Root Cause
Software design
Distribution Pattern
Worldwide distribution. Austria, Denmark, Germany, Singapore, Netherlands, Hong Kong, and Switzerland. US product is not affected.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.