Recall StatusTerminated
Date Initiated2006-03-29
Date Terminated2008-08-08
Product Quantity778
Reason for Recall
The company's initiation of the recall is based on a finding that some units of the Total Knee device may contain faulty pins based in the axis of the knee. Small cracks propagate through the pin cross section until the pin breaks, severely compromising the knee function.
Firm Action
On 3/29/06 a press release was posted on OandP.com, an industry specific web site. An announcement listing involved serial numbers and contact phone numbers was also posted on the company''s web site, ossur.com. A two stage mailing has been constructed. The first letter sent to all customers March 30, 2006, describes the problem and models affected. The second letter, March 31, 2006, is customer specific and lists all of the involved serial numbers by model which were purchased by that customer. Telephone calls will be placed to customers who do not respond to the mailings.
Distribution Pattern
Distribution worldwide in Austria, Canada, Columbia, Hong Kong, Japan, Korea, Kuwait, Malaysia, Mexico, Taiwan, & New Zealand. Involved states are: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, AND WY.
Official Source