Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 127 cm (50"), REF US502, to prevent or treat hypothermia
FDA Recall Z-1198-2025 · Posted 2025-02-20
Advertisement
Recalling FirmAugustine Temperature Management, LLC
Recall StatusOpen, Classified
Product CodeDWJ
Date Initiated2024-12-16
Product Quantity1 unit
Reason for Recall
There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It
turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
Firm Action
Augustine Temperature Management began notifying its consignees on 12/16/2024 via email and visit to exchange the affected product.
Root Cause
Device Design
Distribution Pattern
US Nationwide distribution in the states of TX, AL, NV; AU, NZ.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.