Nucletron Adapter and Transfer Guide Tube, part number AL 13005000, a component used with the VariSource" IX HDR Afterloader. Product Usage: The adapter is part of the VariSource Afterloader System, a computer controlled remote HDR Afterloader used to place a high activity radioactive source within a needle(s) or applicator(s) which have previously been placed for a specific clinical purpose in a patient.

FDA Recall Z-1199-2012 · Posted 2012-03-12

Recall StatusTerminated
Product CodeJAQ
Date Initiated2012-01-05
Date Terminated2012-11-21
Product Quantity122 total units

Reason for Recall

Components of medical device used in brachytherapy may be defective and cause a shifted dose distribution to patients.

Firm Action

Varian Medical Systems sent an Urgent Medical Device Correction - Urgent Field Safety Notice letter dated January 5, 2012 to all affected consignees. The letter identified the affected products, description of problem and recommended user actions to be taken. Consignees were requested to cease use of the affected products and contact Varian BrachyTherapy Helpdesk immediately to arrange return of their affected parts. Consignees were requested to return the enclosed proof of notification form. Consignees were instructed to advise the appropriate personnel working in their radiotherapy department of the content of this letter. For future reference, this document will be posted to the Varian Customer Support website: http://www.MyVarian.com. For further clarification contact your local Varian District or Regional Manager.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution - USA (nationwide) Including the states of CA, FL, IN, KY, MI, MN, MO, OH, TX, VA and the countries of Canada, Japan and Sweden.