6-hole Pubic Symphysis plate (PN 70-0451). The pelvic bone plates for the pelvic ring are a series of plates with varying lengths that function as internal fixation devices for fractures, fusions and osteotomies of the pelvic ring. The plates are strategically pre-contoured where beneficial to the application, and they are secured to the bone with 3.5mm non-locking screws. The Acumed pelvic ring plate is the Pubic Symphysis Plate. The Acumed Pelvic Plating System is intended for use by surgeons with orthopedic training and knowledge of the indications and techniques required for fixation. The device is to be implanted by the surgeon in a sterile operating room setting.

FDA Recall Z-1199-2017 · Posted 2017-02-14

Recalling FirmAcumed LLC
Recall StatusTerminated
Product CodeHRS
Date Initiated2017-01-19
Date Terminated2017-05-08
Product Quantity15 units

Reason for Recall

The 6-hole Pubic Symphysis plate (PN 70-0451) is specified to be made of Commercially Pure Titanium (ASTM F67 - CP Grade 4). However, batch 383813 was manufactured out of Titanium Alloy (Ti-6Al- 4V ELI per ASTM F135 Ti alloy) and was distributed.

Firm Action

Acumed sent an Urgent Medical Device Recall letter dated January 16, 2017, to all affected customers on January 19, 2017. Consignees were instructed to follow special instructions and send survey to [email protected]. Domestic customers with questions or need to return products to Acumed, please contact Acumed Agent Inventory via email [email protected] to obtain a Return Material Authorization (RMA) number or contact your agent directly. International customers with questions or need to return products to Acumed, please contact Acumed Business Service via email at [email protected] or via phone 877-627-9957, choose Option 4 (US and Canada), or call 001-503-627-9957, choose option 4 (outside of US and Canada) have to obtain a Return Material Authorization (RMA) number.

Root Cause

Mixed-up of materials/components

Distribution Pattern

Worldwide Distribution - Costa Rica, Malaysia, Mexico, South Africa, and Spain