Open, Classified

Innova IGS 6, Interventional Fluoroscopic X-Ray System used in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.

FDA Recall Z-1201-2019 · Posted None

Recalling FirmGE Healthcare, LLC
Recall StatusOpen, Classified
Product CodeIZI
Date Initiated2019-03-15
Product Quantity22 devices

Reason for Recall

There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously.

Firm Action

GE Healthcare notified customers on about 03/15/2019 via "URGENT MEDICAL DEVICE CORRECTION" letter. Customer were informed that they can continue to use their Innova IGS 6 or Innova IGS 630 Systems but must follow the following safety instructions: 1. Do not change the size of the field of view from 30cm to 20cm or 20cm to 30cm while releasing the fluoroscopy footswitch pedal. 2. Only perform these 2 actions sequentially. 3. If the system becomes locked up, perform a reset to recover functionalities. Customers were also informed that GE Healthcare will correct all affected products at no cost and a GE Healthcare representative will contact each customer to perform the correction. After the GE representative has updated the system, please destroy previous Innova IGS 6 SW package installation media related to this system (versions prior to IGS6_2.2.1 or IGS6_3.1.1). Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Root Cause

Device Design

Distribution Pattern

Nationwide distribution to AR, CA, CA, FL, MI, NM, NV, SC, TN, TX, VA, WA, WI. International distribution to Algeria, Belgium, China, Colombia, France, Hungary, Italy, Japan, Korea (Republic Of), Kuwait, Libya, Pakistan, Poland, Russia, Saudi Arabia, South Africa, Sudan, Turkey, Vietnam.