da Vinci S. Surgical System, da Vinci Si and da Vinci Si-e Surgical Systems. Intuitive Surgical, Inc. Sunnyvale, CA 94086. The Intuitive Surgical Endoscopic instrument Control System is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.

FDA Recall Z-1202-2012 · Posted 2012-03-13

Recall StatusTerminated
Product CodeNAY
Date Initiated2012-02-06
Date Terminated2012-05-09
Product Quantity1966 systems are in commerce at hospital locations.

Reason for Recall

The holding brake, a component of the da Vinci S Surgical System, may allow passive uncontrolled motion due to gravity during specific power-off conditions.

Firm Action

Intuitive Surgical sent an "URGENT: MEDICAL DEVICE CORRECTION" letter dated February 3, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter provided new instructions for use. Additionally, a Device Correction Response Form was enclosed for customers to complete and return via fax to 408-523-1390. Contact Intuitive Surgical Customer Service at 1-800-876-1310 for questions regarding this notice.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution.