Terminated

Siemens Artis Q and Artis Q biplane systems Product Usage: Interventional Fluoroscopic X-ray

FDA Recall Z-1202-2014 · Posted 2014-03-27

Recall StatusTerminated
Product CodeOWB
Date Initiated2013-12-05
Date Terminated2014-11-06
Product Quantity19

Reason for Recall

There exists the possibility that during clinical operation, a breakdown of the x-ray tube could result in loss of x-ray imaging for the Artis Q and Artis Q biplane systems with a flat detector as 40 HDR (large 30x40).

Firm Action

Siemens sent an Important Customer Safety Notice letter dated November 22, 2013, to affected customers. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to discontinue use until the affected product has been successfully replaced by a Siemen's representative.at no cost.

Root Cause

Device Design

Distribution Pattern

USA Nationwide Distribution in the states of TN, MA, CA, IL, OH, FL, UT, WA, GA, IA, NY, MO, IN, and CO.