Siemens Digital Linear Accelerators of type: ARTISTE MV System MEVATRON M2/Primus Mid-Energy PRIMUS PRIMUS HI ONCOR Impression ONCOR Impression Plus ONCOR Avant Garde ONCOR Expression ARTISTE, ONCOR, and PRIMUS running the following: Control Console from software version 13.0.302 and higher; Control Console from software version 9.2.400 and higher; and Control Console from software version 11.0.400 and higher.
FDA Recall Z-1204-2019 · Posted None
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeIYE
Date Initiated2019-03-20
Date Terminated2020-06-18
Product Quantity120 US units
Reason for Recall
Advisory issued not to disable or bypass the AMP function. Disabling or bypassing the AMP function for reasons of comfort is not advisable may lead to patient or user injury due to collision with LINAC equipment disabling the Auto Field Sequence (AFS) Automatic Motion Protection (AMP) function implemented at the control of the Digital Linear Accelerator.
Firm Action
Siemens Healthcare issued CUSTOMER SAFETY ADVISORY NOTICE to accounts via electronic mail addresses and/or USPS certified mail on March 20. Letter identifies the issue , action to take: due to potential for collisions of the LINAC equipment with patients, Siemens Healthcare GmbH advises
customers to not to disable or bypass the AMP function. Customers who decide to disable or bypass the AMP function, do so at their own responsibility, duly weighing patient benefit and risk. Disabling or bypassing for reasons of comfort is not advisable. Notify and instruct accordingly all the staff at your organization who need to be aware of this notice. In addition, if you may have further distributed this product, please identify your customers and notify them. Only in the event you choose to disable or bypass the AMP Function, Siemens requests that you carefully read and complete the following form. Please sign, scan, and email the completed form mail to:[email protected]
Root Cause
Use error
Distribution Pattern
Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K121295ARTISTETM SOLUTION WITH SYS VCIOSiemens Medical Solutions USA, Inc. · 2012-06-27K072485ARTISTE MV SASiemens Medical Solutions USA, Inc. · 2007-12-27K060226ONCOR EXPRESSIONSiemens Medical Solutions USA, Inc. · 2006-03-15K031764ONCOR AVANT-GARDE WITH COHERENCE WORKSPACESSiemens Medical Solutions USA, Inc. · 2003-09-05K993425PRIMUSSiemens Medical Solutions USA, Inc. · 1999-11-26