RT Elements Software revisions of the RT Elements applications have a specific software version number. Specifically the following RT Elements applications/versions are affected: - Cranial SRS 1.0.0 and 1.5.0 - Spine SRS 1.0.0 and 1.5.0 - Multiple Brain Mets SRS 1.5.0 - RT QA 1.0.0 and 1.5.0 Product Usage: The RT Elements are applications for radiation treatment planning for use in stereotactic, conformal, computer planned, Linac based radiation treatment of cranial, head and neck and extracranial lesions.

FDA Recall Z-1205-2019 · Posted None

Recalling FirmBrainlab AG
Recall StatusTerminated
Product CodeMUJ
Date Initiated2019-03-11
Date Terminated2021-05-12
Product QuantityUS - 164 units; Outside the US - 333

Reason for Recall

There is a potential for an incorrect dose distribution calculation by Brainlab RT Elements software (for affected versions) under specific circumstances when using the Pencil Beam algorithm on the GPU (graphics card), as is the default system setting.

Firm Action

On March 11, 2019, Brainlab sent FIELD SAFETY NOTICE/PRODUCT NOTIFICATION emails to their consignees advising them of the problem. Brainlab will disable the Pencil Beam GPU calculation in the settings of the affected RT Elements Radiation Treatment Planning Software, and thus correct the described issue permanently for all affected customers.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution - US Nationwide in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IN, KY, LA, MA, MD, MS, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, & WV. Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, El Salvador, Finland, France, Germany, Hong Kong, Iran, Ireland, Israel, Italy, Japan, Jordan, Lebanon, Martinique, Mexico, Monaco, Morocco, Netherlands, Poland, Portugal, Russian Fed., South Africa, South Korea, Spain, Switzerland, Taiwan, Tunisia, United Arab Emirates, and United Kingdom.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.