Software Versions V1.2R,or earlier for the Harvard 2, Dual Infusion Pump (P/N 2002-001)
FDA Recall Z-1208-03 · Posted 2003-09-11
Advertisement
Recalling FirmHarvard Clinical Technology
Recall StatusTerminated
Product CodeFRN
Date Initiated2003-08-11
Date Terminated2012-05-07
Product Quantity384 units
Reason for Recall
Under specific sequence conditions, the unit may apper to be infusing, but is not
Firm Action
Harvard Medical Technology inc. notified customers by letter on 8/11/03, advising users of the software upgrade to version V1.3T. Facilities will be visited by technicians who will upgrade the software.
Root Cause
Other
Distribution Pattern
Nationwide