Terminated

VNVD Veronikis Neo Vaginal Dilator - Complete Set Catalog No. 270-100, VNVD Veronikis Neo Vaginal Dilator - 1.5cm Tip:Width:0.625in Catalog No. 270-115 (sold individually), VNVD Veronikis Neo Vaginal Dilator - 2.5cm Tip:Width:0.625in Catalog No. 270-125 (sold individually), VNVD Veronikis Neo Vaginal Dilator - 3.5cm Tip:Width:0.625in Catalog No. 270-135 (sold individually), VNVD Veronikis Neo Vaginal Dilator - 4.5cm Tip:Width:0.625in Catalog No. 270-145 (sold individually), VNVD Veronikis Neo Vaginal Dilator - 5.5cm Tip:Width:0.625in Catalog No. 270-155 (sold individually), VNVD Veronikis Neo Vaginal Dilator - Partial Taper Couple: Width:0.625in - 1in Catalog No. 270-165 (sold individually), VNVD Veronikis Neo Vaginal Dilator - Full Taper Couple: Width:0.625in - 1in Catalog No. 270-175 (sold individually), VNVD Veronikis Neo Vaginal Dilator - Coupler 3cm Length: Width: 1in Catalog No. 270-185 (sold individually), VNVD Veronikis Neo Vaginal Dilator - Small Base, Screw Length 2cm Tip: Width: 1in Catalog No. 270-195 (sold individually), VNVD Veronikis Neo Vaginal Dilator - 1.5cn Tip: Width:0.625in Catalog No. 270-199 (sold individually). For use in securing open the opening during examination of the vaginal body cavity. For the Non-Surgical Neo Vagina.

FDA Recall Z-1208-2010 · Posted 2010-03-25

Recall StatusTerminated
Product CodeHDQ
Date Initiated2009-04-09
Date Terminated2010-09-03
Product Quantity22

Reason for Recall

Marina Medical, Sunrise, FL initiated recall of 46 products of 350 series devices and 47 products of 270 series products because the material used to manufacture the products was not proven to be biocompatible, resulting in risk of temporary skin or tissue irritation.

Firm Action

A "PRODUCT RECALL NOTICE" letter dated April 8, 2009, was sent to the customers. The letter describes the product, problem and the action to be taken by the customers. The customers are asked to identify the subject part in their instrument inventory and quarantine it. The Quality Control team at Marina Medical will contact the customer with instructions to return the part. When contacted, the customer should report any incidents of adverse reactions that have been observed at their institution. If you have any questions, please call 1-800-697-1119 ore (954) 924-4418.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide distribution.