Mar Cor Purification Millenium HX Portable Reverse Osmosis System. Model 3025975. Product Usage: Water Purification System is intended to be used as a dialysis accessory to produce water through reverse osmosis for use with hemodialysis equipment..

FDA Recall Z-1213-2014 · Posted 2014-03-14

Recalling FirmMar Cor Purification
Recall StatusTerminated
Product CodeFIP
Date Initiated2013-06-26
Date Terminated2015-01-22
Product Quantity232 units

Reason for Recall

High inlet water pressure beyond specification and the solenoid valve does not meet application which could result in a delay in treatment.

Firm Action

Mar Cor sent an Urgent Medical Device Correction letter dated March 5, 2014 to all affected customers. The letter identifies the affected devices, what the issue is and tells customers to contact Mar Cor to schedule correction. A response form was asked to be completed and returned. For questions contact Scott Nichols, Project Coordinator/Service Trainer at 1-800-633-3080.

Root Cause

Component design/selection

Distribution Pattern

USA Nationwide Distribution including Puerto Rico.