Sonimage HS1 Ultrasound Kit AC adapter, Product Number: 1. A9RP, UDI: 04560141945463; 2. AAEU, UDI: 04560141946385; 3. A9RR, UDI: 04560141945470

FDA Recall Z-1214-2018 · Posted None

Recall StatusTerminated
Product CodeIYN
Date Initiated2017-12-04
Date Terminated2019-10-28
Product Quantity30

Reason for Recall

There is a risk of circuit board breakage within the main housing of the AC Adapter unit. Smoke may be emitted from the AC Adapter as a direct result.

Firm Action

Konica Minolta is preparing for the final AC Adapter replacement. Upon receipt of the new AC Adapter, owners of SONIMAGE HSl System will be asked to replace the existing AC Adapter. Additionally, instructions for the return of the current AC Adapter will be provided.

Root Cause

Device Design

Distribution Pattern

Distributed in the following U.S. states: AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, WA, WI and the District of Columbia. Distributed in Canada and Brazil.