Terminated

TriMed Hex Cortical Screw 4.0, 40 mm, REF HEX4.0-40, non-Sterile, RX, a component of the TriMed Ankle Fixation System. The HEX4.0 40 bone screw is a component of the TriMed Ankle Fixation System. The HEX4.0-40 Screw is a single piece machined cortical bone screw. The HEX4.0-40 screw is employed in combination with a TriMed Ankle Bone Plate such as a Sidewinder Plate, Ankle Hook Plate, Cluster, Semi-Tubular Plate, or the Locking Plate. Use of these implants is not a substitute for normal tissue healing. The TriMed Ankle Bone Plates and Screws are designed to provide additional constraint of movement of a fractured bone and are intended only as an aid to fix the fracture in place during the healing process.

FDA Recall Z-1214-2019 · Posted None

Recalling FirmTriMed Inc.
Recall StatusTerminated
Product CodeHRS
Date Initiated2018-08-24
Date Terminated2020-12-28
Product Quantity13 screws

Reason for Recall

The bone screw is shorter than the intended design specification.

Firm Action

The recalling firm, TriMed, issued a "URGENT MEDICAL DEVICE RECALL" letter via email on 8/24/2018 to the distributors and third-party distributor. A follow-up letter was issued 3/28/2019 requesting downstream customers be notified to the user level. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: " If the customer further distributed Hex Cortical Screws, 4.0mm that is the subject of this recall, to forward the letter and the return response card to those entities that utilized the impacted product. " If the customer utilized the Hex Cortical Screws, 4.0mm that is the subject of this recall, then the customer is to: o Review customer inventory (including any kits that you have prepared for surgical operation) to see if they have any of the impacted product. o If a customer does have impacted product within its inventory, remove the product from use. o Review customer records to see if the impacted product was implanted. If the customer has records that impacted product was implanted, the customer should apply medical judgment to determine whether additional patient monitoring activities should be performed to evaluate implant placement, implant structural integrity, and fracture union. o Return to TriMed a return response card by 11 April 2019. Please email the response to: [email protected]. Please contact TriMed via telephone or email as per below if you have any questions: Telephone*: 800-633-7221 or Email: [email protected] * Hours of operation: 8:00AM to 5:00PM PST (Monday through Friday, excluding holidays).

Root Cause

Component design/selection

Distribution Pattern

US Distribution to states of FL and MO.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑