Boston Scientific IQ Guide Wire. Catalog No. 38950-020, 300 cm, Straight Tip, Single Pack. Catalog No. 38950-022, 300 cm, Straight Tip, 5 pack. Sterile EO. Manufactured for: Boston Scientific Corporation, One Boston Scientific Place, Natick, MN 01760 USA. Boston Scientific Guide Wires are steerable guide wires available in a variety of lengths and diameters. The distal tip is shapeable, or, alternatively, is available in a preshaped ''J'' tip.

FDA Recall Z-1215-06 · Posted 2006-07-12

Recalling FirmBoston Scientific
Recall StatusTerminated
Product CodeDQX
Date Initiated2006-05-30
Date Terminated2007-06-28
Product Quantity1455 units total

Reason for Recall

Boston Scientific discovered through an internal inspection process that the PTFE (polytetrafluoroethylene) coating is not consistently adhering to the transition area of the distal end of the guide wires which may cause PTFE to flake off.

Firm Action

On May 30, 2006, Boston Scientific notified customers of the IQ Guide Wire manufactured from February 3, 2006, through March 15, 2006. BSC requested a reply whether or not they still have product in their institution. Replacements will be issued for returned product. UPDATE: A second recall notification was sent to affected customers 07/07/06. This second notification included Inner Pouch Number (UPN) and Inner Pouch Lot/Batch #.

Root Cause

Other

Distribution Pattern

Nationwide (AR, AZ, FL, GA, IA, ID, IL, IN, KY, LA, MA, MI, MN, MO, NC, NH, NJ, NM, NV, OH, OK, PA, TN, TX, UT, VA, WI, WV) and Puerto Rico. Chile, Germany, Taiwan.