Terminated

The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase (ALTI and ALTI) and Aspartate Aminotransferase (AST) methods on the Dimension Vista System. For prescription use only.

FDA Recall Z-1216-2017 · Posted None

Recall StatusTerminated
Product CodeJIX
Date Initiated2016-05-05
Date Terminated2018-02-02
Product Quantity13,147 units

Reason for Recall

Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.

Firm Action

Siemens mailed an Urgent Medical Device Notification letter to all affected customers on May 10, 2016. The notice provided information on actions to be taken by the customer to work around the issue. For further questions, please call (800) 441-9250.

Root Cause

Under Investigation by firm

Distribution Pattern

Distributed throughout the United States

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.