Misys Laboratory System versions 5.2 through 5.3.2 using LabAccess Results

FDA Recall Z-1217-03 · Posted 2003-09-11

Recall StatusTerminated
Product CodeJQP
Date Initiated2003-07-22
Date Terminated2004-06-25
Product Quantity171

Reason for Recall

Software design deficiency. If the dilution factor is added to the same result twice, incorrect specimen values are reported.

Firm Action

Product Safety Notice, PSN-03-L04 was issued to all affected customers by fax on 07/22/2003. This included a procedural workaround to use until a software correction is made available or to use indefinitly at the user''s discretion. Completion of the recall , including release of the software fix is expected in December 2003.

Root Cause

Other

Distribution Pattern

Nationwide and to United Kingdom, Canada, Denmark, Ireland, Saudi Arabia, United Arab Emirates.