Prevent G Safety Winged Blood Collection Set Glide - Rx Only -Sterile/EO - Product Usage: The IMPROSAFE Blood Collection Set is single-use, sterile, winged needle bonded to a flexible tubing with luer adapter. The needle consists of (1) a winged needle, (2) a flexible tubing, (3) a luer adaptor. (4) a safety shield, The products and evacuated blood collection tube/ syringe are used together for the collection of venous blood.

FDA Recall Z-1217-2019 · Posted None

Recall StatusOpen, Classified
Product CodeFMI
Date Initiated2017-09-22
Product Quantity550,000

Reason for Recall

Complain due to the hub needle came out when the nurse draw the blood and removed the tube.

Firm Action

Several e-mails dated May 2017 thru June 2017between the manufacturer and the sole distributor for this lot are documented.

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide distribution in the state of Kansas.