Vanguard Total Knee, PUNCH THRU TRL PLATES, 63 MM Item No: 32-487261, 67 MM Item No: 32-487262, 71 MM Item No: 32-487263, 75 MM Item No: 32-487264, and 79 MM Item No: 32-487265
FDA Recall Z-1219-2017 · Posted None
Advertisement
Recalling FirmZimmer Biomet, Inc.
Recall StatusTerminated
Product CodeJWH
Date Initiated2016-12-15
Date Terminated2018-06-29
Product Quantity26
Reason for Recall
Zimmer Biomet is conducting a medical device recall for ARCOS & TPRLC broaches and rasps due to potential alumina inclusions in the raw material batch used to produce the affected products. Inclusions contained within the finished product could lead to the cracking and separation of the instrument.
Firm Action
Zimmer Biomet is conducting a medical device recall for ARCOS & TPRLC broaches and rasps due to potential alumina inclusions in the raw material batch used to produce the affected products. Inclusions contained within the finished product could lead to the cracking and separation of the instrument
Root Cause
Material/Component Contamination
Distribution Pattern
Domestic: 0 Foreign: Singapore & Malaysia