Terminated

AutoDELFIA Neonatal immunoreactive trypsin(ogen) (IRT) kit, intended for the quantitative determination of human immunoreactive trypsin(ogen) (IRT) in blood specimens. (UDI: (01)06438147298716(17)190430(10)656479); Product Number: B005-212

FDA Recall Z-1220-2018 · Posted None

Recall StatusTerminated
Product CodeJNO
Date Initiated2017-09-25
Date Terminated2018-09-06
Product Quantity240

Reason for Recall

The Quality Control Certificate for certain AutoDELFIA Neonatal IRT Kits is missing information required to assess the acceptability of measured control values which may result in a delay in reporting IRT results.

Firm Action

Actions to be taken: Inspect your inventory of AutoDELFIA Neonatal IRT kits for batch number 1065647901 and replace the Quality Control Certificate included in the kit packaging with the revised certificate attached to this letter. If you have already used the kit lot there is no need to retest the samples reported if the results of the controls included in the assay were within your established acceptance range. Complete and return the enclosed response form identifying the quantity of affected kits in your possession. Please inform others affected in your organization accordingly. To comply with regulatory requirements we require that you complete the enclosed Response Form and return it.

Root Cause

Process control

Distribution Pattern

Distributed to CT and TX.