Terminated

St. Jude Medical Confirm Rx Insertable Cardiac Monitor, REF DM3500, Sterile. Product Usage: The St. Jude Medical Confirm Rx insertable cardiac monitor (ICM) is designed to detect arrhythmias and wirelessly transmit data to the Merlin.net Patient Care Network (PCN). It is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias. It is also indicated for patients who have been previously diagnosed with atrial fibrillation or who are susceptible to developing atrial fibrillation.

FDA Recall Z-1220-2019 · Posted 2019-04-25

Recalling FirmAbbott
Recall StatusTerminated
Product CodeMXC
Date Initiated2018-11-19
Date Terminated2021-04-28
Product Quantity32 devices

Reason for Recall

The device is unable to pair with the mobile app due to the device incorrectly determining the certificate has expired.

Firm Action

Abbott sent an Important Medical Device Correction letter dated November 20, 2018 to affected customers. The letter identified the affected product, problem and actions to be taken. Should you have questions about this issue, contact your local Abbott Representative or Abbott Support at 1-800-727-7846 (Opt3) (U.S.), 8:30am - 5:30pm Central Time Monday thru Friday.

Root Cause

Software design

Distribution Pattern

US Nationwide Distribution in the states of AZ, CA, CO, ID, IL, LA, MA, MI, MO, NE, NH, NJ, NY, TN, TX, and WI.