Terminated

VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658 Product Usage: The VITROS XT 7600 Integrated System is intended for use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides, VITROS Chemistry Products MicroTip Reagents, and VITROS Immunodiagnostic Products Reagents

FDA Recall Z-1222-2019 · Posted None

Recall StatusTerminated
Product CodeJJE
Date Initiated2019-03-11
Date Terminated2020-12-07
Product Quantity80

Reason for Recall

One of the software algorithms used to detect sample dispense errors was inadvertently disabled. Because of this, sample dispense errors may lead to incorrect results being reported without an error code to alert the user.

Firm Action

Urgent Product Correction Notification letters dated 3/11/19 were sent to customers.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution - US Nationwide in the states of: AZ, CA, FL, IL, OH, TX, and VA. The products were distributed to the following foreign countries: Australia, Belgium, Canada, Chile, Columbia, France, India, Italy, Japan, Portugal, Singapore, Spain, United Kingdom