Green Spec Fiberoptic Laryngoscope Handle - Stubby/Short; a battery operated fiberoptic laryngoscope handle for use with green coded fiberoptic systems and single use laryngoscope blades; Teleflex Medical, Bannockburn, IL 60015; catalog number 004413300
FDA Recall Z-1225-06 · Posted 2006-07-13
Advertisement
Recalling FirmTeleflex Medical
Recall StatusTerminated
Product CodeCCW
Date Initiated2006-06-15
Date Terminated2009-08-28
Product Quantity1,408 laryngoscopes
Reason for Recall
Teleflex Medical has identified that the product may malfunction, causing the handle to heat up. There is a potential for the heated handle to burn the user.
Firm Action
Teleflex Medical sent Urgent Medical Device Recall letters dated 6/15/06 to the direct accounts, informing them of the potential for the handle to heat up, and requesting that they cease use of the affected units and return them to Teleflex Medical for replacement. The accounts were instructed to complete and return by fax the enclosed Acknowledgement & Stock Status Form, indicating the number of units being returned for replacement. Any questions were directed to Linda Todd at 1-800-334-9751, ext. 4951.
Root Cause
Component change control
Distribution Pattern
Nationwide and Canada, Colombia, Costa Rica and Chile.