Terminated

ReliOn 0.5cc 30 g x 8mm Blister Pak U-100 syringe 30 gauge x 5/16" length, for subcutaneous injection of insulin.

FDA Recall Z-1226-2007 · Posted 2007-09-20

Recall StatusTerminated
Product CodeFMI
Date Initiated2007-06-21
Date Terminated2008-06-03
Product Quantity644,500 units

Reason for Recall

Complaints received indicating open seals on blister paks, potentially compromising the sterility of the product.

Firm Action

Urgent Product Recall notifications were sent on 6/21/07 asking them to discontinue use of product, remove from inventory and complete Tracking Verification form.

Root Cause

Other

Distribution Pattern

Nationwide.