INTELECT ADVANCED, Model No. 2762CC, Chattanooga Group, Hixson, TN 37343, Made in the USA, Not for sale in the U.S.A.
FDA Recall Z-1227-2008 · Posted 2008-06-25
Advertisement
Recalling FirmChattanooga Corp Inc
Recall StatusTerminated
Product CodeMPH
Date Initiated2007-12-17
Date Terminated2010-02-19
Product Quantity2993 units
Reason for Recall
Device malfunction: causing electric shock and burn.
Firm Action
Beginning on 12/17/2007, Chattanooga Corp. sent two recall notification letters, via UPS mail, to its consignees informing them of the situation. The letters stated that the recall was to the user level and requested that the user perform the software upgrade, which will eliminate the possibility of shock and burn.
Root Cause
Device Design
Distribution Pattern
Worldwide Distribution - including USA
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.