Terminated

Covidien Duet TRS Articulating 60- 3.5MM, Single Use Loading Unit Ref: DUET6035A The ENDO GIA Universal Staplers with DUET TR S Loading Units have applications in abdominal, gynecologic and pediatric surgery for resection, transection, and creation of anastomosis. They may be used for transaction and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas

FDA Recall Z-1227-2012 · Posted 2012-03-19

Recalling FirmCovidien LP
Recall StatusTerminated
Product CodeGDW
Date Initiated2012-02-16
Date Terminated2021-03-04
Product Quantity1158 units

Reason for Recall

Missing component results in the stapler not firing.

Firm Action

The firm, Covidien, sent a "Medical Device Recall" letter dated February 16, 2012 to customers via federal express. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue use of Lot# N1H003LX of Duet TRS Articulating Loading Unit and segregate affected product; return inventory to Field Returns Department, 195 McDermott Road, North Haven, CT 06473 and contact customer service at [email protected] or (800) 722-8772, option #1 for RGA # prior to returning; and complete and return the Recall Product Return Form via fax to (203) 822-6009 (all affected units must be returned with completed form through the Distributor). If you have any questions or concerns, please do not hesitate to contact your Covidien representative or Quality Assurance at 203-492-5232, or Customer Service at 800-722-877 option 1 customer service.

Root Cause

Process control

Distribution Pattern

Worldwide distribution: USA (nationwide) and country of: Venezuela.