Open, Classified

BIOPHEN Protein C 5 kit is for in vitro quantitative determination of Protein C activity on human citrated plasma. Protein C is a glycoprotein, vitamin K dependent, which inhibits coagulation. Protein C in plasma is measured following a specific activation with Protac (Agkistrodom Contortrix), an enzyme extracted from snake venom. Activated protein C hydrolysis the chromogenic substrate (SaPC-21) which release para-nitroaniline (pNA). The amount of pNA released (measured by absorbance at 405 nm) is directly proportional to the concentration of Protein C in the specimen.

FDA Recall Z-1227-2023 · Posted 2023-03-06

Recall StatusOpen, Classified
Product CodeGGP
Date Initiated2023-02-01
Product Quantity2 units

Reason for Recall

New warnings and precautionary statements on product labelling: Reagent 2 (R2) of BIOPHEN Protein C 5 contains the chemical substance - Cesium Chloride which changed classification and has become hazardous according to the CLP regulation ((EC) No 1272/2008) is suspected of damaging fertility

Firm Action

Aniara issued Medical Field Correction Letter (FSCA#29) prepared by HYPHEN BioMed and emailed to the customer by Aniara Diagnostica on 2/1/23. Letter states reason for recall, health risk and action to take: New labelling will be implemented on all new lots as follow: x A new pictogram added on reagent and box label: GHS08 x Signal Word: Warning x Hazard Statements: H361f - Suspected of damaging fertility x Precautionary Statements: o P201 : Obtain special instructions before use. o P280 : Wear protective gloves/protective clothing/eye protection/face protection. o P308 + P313 : IF exposed or concerned: Get medical advice/attention. o P405 : Store locked up. o P501 : Dispose of contents/container in consultation with your regional waste disposer x Hazardous ingredients : Cesium chloride The IFU, label and SDS will be updated. Please return completed and signed Form to your distributor.

Root Cause

Other

Distribution Pattern

US Nationwide distribution in the state of NC.