Synchron CX Immuno-Protein Cal, part number: 442840, other product code: CKA The Beckman Coulter SYNCHRON CX Immune Protein Calibrator Set, in conjunction with SYNCHRON Immuno Protein Reagents, is intended for use on SYNCHRON CX Systems for the calibration of IgA, IgG, IgM, and Transferrin (TRF).
FDA Recall Z-1229-2012 · Posted 2012-03-20
Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeJIT
Date Initiated2009-08-26
Date Terminated2012-04-17
Product Quantity1,923 units total (1,882 units in US)
Reason for Recall
A recall was initiated because Beckman has confirmed that the standardization of the Immuno Protein Calibrator Set has been re-established to IFCC Reference Preparation for Plasma Proteins, BCR-470 beginning with lot number M901330.
Firm Action
Beckman Coulter sent an Urgent Field Safety Notice letter dated August 24, 2009, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to share the information with their laboratory staff and retain the notification as part of their laboratory Quality System documentation. Customers with questions regarding the product correction action were instructed to contact their local Beckman Coulter Representative.
For questions regarding this recall call 714-993-8916.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide Distribution - USA (nationwide) and Canada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.