PRISMAFLEX HF1000 SET, Product Code 107142; Dialyzer, High Permeability With Or Without Sealed Dialysate System

FDA Recall Z-1230-2026 · Posted 2026-02-02

Recall StatusOpen, Classified
Product CodeKDI
Date Initiated2026-01-06
Product Quantity115,112 units

Reason for Recall

There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.

Firm Action

Vantive issued an Urgent Medical Device Correction notice to its consignees on 01/06/2026 via USPS first class mail. The notice explained the issue, hazard involved, and requested the following: 1. You may continue to perform therapy using the Prismaflex sets listed above. Please monitor the deaeration chamber during therapy to ensure it remains in an upright position within the holder. 2. Vantive is aware of customers who have experienced this issue and attempted to further secure the deaeration chamber in an upright position. If attempting to secure the deaeration chamber in an upright position, please consider the following: . Ensure there are no kinks in the tubing and that the deaeration chamber remains visible. . If the Prismaflex Control Unit issues an Air in Blood alarm, check if air is present. If there is no air present, check if the deaeration chamber is dislodged from the holder and ensure there are no clots present. 1. If no dislodgement, please follow the operator s manual instructions for Air in Blood alarm. 2. In case of dislodgement with clotting, follow the associated instructions within the Prismaflex Control Unit operator s manual to stop the therapy and replace the set. 3. Only in case of dislodgement without clotting or any other alarms, if the chamber is secured in an upright position, follow the associated instructions within the Prismaflex Control Unit operator s manual to proceed with treatment. Facilities who further distributed the products were directed to notify those departments or facilities. Dealers, wholesalers, distributor/resellers, or original equipment manufacturers (OEM) who further distributed the products were directed to notify their customers. For general questions regarding this communication, please contact Vantive Customer Care at 833-542-CRRT (2778). Vantive Customer Care is available 24 hours a day, seven days a week.

Root Cause

Component design/selection

Distribution Pattern

US Nationwide distribution.