Triathlon X3 Total Stabilizer + Tibial Insert For use with Triathlon Universal Baseplates, knee prosthesis. #5 13mm Stryker Orthopaedics

FDA Recall Z-1231-2007 · Posted 2007-09-25

Recall StatusTerminated
Product CodeJWH
Date Initiated2007-07-31
Date Terminated2007-12-05
Product Quantity14

Reason for Recall

Stabilizer posts, packaged with the Triathlon TS Plus Tibial inserts have a specification anomaly related to length, resulting in a potential for the post to slightly protrude above the top surface of the plastic post.

Firm Action

Recall notification letters and product acknowledgment forms were send via Federal Express on August 2, 2007.

Root Cause

Other

Distribution Pattern

Product distributed to 5 hospitals in GA, OH, MN, TX.