Terminated

Clinical Chemistry Lipase, List Numbers: 7D80-20 and 7D80-30

FDA Recall Z-1232-2007 · Posted 2007-09-18

Recall StatusTerminated
Product CodeCHI
Date Initiated2007-06-01
Date Terminated2012-04-05
Product Quantity7260

Reason for Recall

Lipase results are falsely elevated when a MULTIGENT Acetaminophen test is pipetted immediately before a Lipase test from the same sample cup or tube on the ARCHITECT c8000.

Firm Action

The firm initiated the recall on 06/01/2007 by sending recall letters with attached Customer Reply form via FedEx to each direct account that received the recalled product. Customers were informed of the issue and given instructions for modification to the SmartWash feature in the Lipase Assay Configuration in order to prevent Acetaminophen from contaminating Lipase.

Root Cause

Other

Distribution Pattern

Worldwide, including USA, Australia, Brazil, Canada, Chile, El Salvador, Dominican Republic, Ecuador, Germany, Hong Kong, Japan, Mexico, New Zealand, Panama, Puerto Rico, Singapore, South Korea, Thailand, and Venezuela.