Terminated

Dimension Enzymatic Creatinine (EZCR) reagent, SMN 10471520

FDA Recall Z-1232-2020 · Posted 2020-01-24

Recall StatusTerminated
Product CodeJFY
Date Initiated2019-12-18
Date Terminated2020-06-25
Product Quantity2,198 units

Reason for Recall

There is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.

Firm Action

Siemens Healthcare Diagnostics initiated a customer communication for this issue via letter beginning December 18, 2019.

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide

Related FDA 510(k) Records