Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

ENDOBON XENOGRAFT GRANULES 0.5 ML (Bovine bone substitute) Endobon is a bone substitute of bovine origin formed from ceramised bone for implantation or transplantation when bone is lost due to trauma or requested during reconstructive surgery.

FDA Recall Z-1233-2014 · Posted 2014-03-19

Recalling FirmBiomet 3i, LLC
Recall StatusTerminated
Product CodeLYC
Date Initiated2013-10-03
Date Terminated2014-12-11
Product Quantity704

Reason for Recall

Individual pieces of a batch of Biomet's Endobon Xenograft Granules 0.5 ML (Bovine Bone substitute) may have an incorrect label with the wrong batch number.

Firm Action

Biomet 3i sent a Medical Advisory Notice dated October 3, 2013 via mail, facsimile or telephone to all affected customers. The notice identified the product, the problem, and the action to be taken by the customers. The customers were instructed to do the following: 1) Review their inventor for any incorrectly labeled products and return them to their local Biomet distributor. 2) Provide this information to persons in their organization that uses or order these products. 3) Ensure that a copy of this letter is provided to any other organization to which the products maybe have been transferred. Customer were also instructed to respond to the attached Response Form via fax at 561-514-6316 or email to: [email protected]. For any questions or concerns, call Biomet 3i Customer Service at 1-800-342-5454. For questions regarding this recall call 561-776-6706.

Root Cause

Error in labeling

Distribution Pattern

Worldwide Distribution - USA including CA, DC, FL, LA, MA, MO, NC, NJ, NY, TX, VA, WA and Hawaii and Internatially to Austria, Belgium, Canada, Switzerland, Germany, Spain, France, Great Britain (UK), Ireland, Mexico, Poland, Portugal, and Singapore.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑