Recall StatusTerminated
Date Initiated2015-02-05
Date Terminated2015-04-10
Product Quantity3109 - total - all model numbers
Reason for Recall
The manufacturers seal may not be present on the Nylon pouch resulting in lack of assurance for sterility.
Firm Action
Zipline sent an Urgent Voluntary product Recall letters to customers on February 5, 2015 by e-mail with possible follow up on site by firm representatives. The letter identified the affected product, problem and actions to be taken. Customers are asked to quarantine inventory immediately and examine the affected lots for complete pouch seals. Any product found with an incomplete seal should be destroyed or returned to ZipLine Medical. Please complete the enclosed response form as soon as possible, even if you no longer have stock of the lot numbers subject to this correction. This information is required to reconcile the inventory. Please contact Customer Service at 1-888-326-8999 to coordinate any product returns, replacement stock, or credit notes.
Root Cause
Packaging process control
Distribution Pattern
Worldwide Distribution - US Nationwide and the countries of Panama, Tehran, Ireland, Sweden, Japan, Hong Kong, Finland, Pakistan, Taiwan, Malaysia.