Sensus Healthcare Inc, SRT-100 Vision IPX 0, HFUS Module IPX 1, with software version 1.9.2; Superficial X-ray Radiation Therapy System with Ultrasonic Imaging Capabilities
FDA Recall Z-1235-2023 · Posted 2023-03-08
Advertisement
Recalling FirmSensus Healthcare, Inc.
Recall StatusTerminated
Product CodeJAD
Date Initiated2023-02-10
Date Terminated2024-01-31
Product Quantity23 units
Reason for Recall
When the user pauses the beam delivery, the dose timer is reset (to zero), and as a result may deliver an unexpected dose to the patient.
Firm Action
Sensus healthcare issued an "Urgent Device Correction" to its consignees on 02/10/2023 by email. The notice requested the following actions:
"Do not initiate beam delivery after a pause & resume sequence without verifying the amount of dose delivered.
Do evaluate any patient treatments with version 1.9.2 that followed the pause & resume
workflow.
Do verify that the amount of dose to be delivered is as expected after each pause & resume
sequence."
The firm has developed a permanent software resolution for this issue. Customer service will contact the customer to coordinate upgrading to this revised software.
Questions: Contact Sensus Customer Support at [email protected] or 561-578-4982
Root Cause
Software design
Distribution Pattern
US Distribution to states of: RX, IN, NM, FL, NY, FA, NC, IA, and IL.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.