3 Series Full Body Phototherapy Device with Smart Touch Control System, Model #SNBSM-0048.
FDA Recall Z-1236-04 · Posted 2004-07-30
Advertisement
Recalling FirmDaavlin Distributing Company
Recall StatusTerminated
Product CodeFTC
Date Initiated2003-09-01
Date Terminated2007-07-24
Product Quantity23 units distributed domestically and 23 units distributed to foreign accounts.
Reason for Recall
Software errors cause the unit to not deliver the prescribed therapy and do not allow the patient records to be saved.
Firm Action
The firm has software patches to correct the programming problems as they are reported by the users. These patches are being installed either by a service representative for the manufacturer or by the customer
Root Cause
Other
Distribution Pattern
Nation wide plus Canada, the Philippines, Saudi Arabia, Turkey & Germany