Axiom Artis Systems operating with A100 Generator
FDA Recall Z-1237-2010 · Posted 2010-04-05
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeMQB
Date Initiated2010-03-04
Date Terminated2012-07-05
Product Quantity1373 units
Reason for Recall
Due to a contact problem at the low voltage power supply of the generator controller, in rare cases where a user has invoked radiation, it may occur that no radiation is released.
Under certain circumstances, radiation is possible by repeated release of the x-ray footswitch or button.
To remedy the issue described above, a revised cabling has been installed in all potentially affected syste
Firm Action
The recalling firm issued a Customer Safety Advisory to affected customers via Update Instructions AX006/10/S. The letter informs customers of the potential issues and provides additional information on the implemented resolutions. For additional information, please contact the firm at 888-826-9702.
Root Cause
Nonconforming Material/Component
Distribution Pattern
The product was shipped to medical facilities nationwide.