Axiom Artis Systems operating with A100 Generator

FDA Recall Z-1237-2010 · Posted 2010-04-05

Recall StatusTerminated
Product CodeMQB
Date Initiated2010-03-04
Date Terminated2012-07-05
Product Quantity1373 units

Reason for Recall

Due to a contact problem at the low voltage power supply of the generator controller, in rare cases where a user has invoked radiation, it may occur that no radiation is released. Under certain circumstances, radiation is possible by repeated release of the x-ray footswitch or button. To remedy the issue described above, a revised cabling has been installed in all potentially affected syste

Firm Action

The recalling firm issued a Customer Safety Advisory to affected customers via Update Instructions AX006/10/S. The letter informs customers of the potential issues and provides additional information on the implemented resolutions. For additional information, please contact the firm at 888-826-9702.

Root Cause

Nonconforming Material/Component

Distribution Pattern

The product was shipped to medical facilities nationwide.