Calculator/ Data Processing Module for Clinical Use
FDA Recall Z-1240-03 · Posted 2003-10-15
Advertisement
Recalling FirmMisys Healthcare Systems
Recall StatusTerminated
Product CodeJQP
Date Initiated2003-07-23
Date Terminated2004-04-23
Product Quantity466
Reason for Recall
Software Defect. Data used in the diagnosis and or treatment of a patient's illness are missing from the patient result report.
Firm Action
Notification to all affected customers was sent by fax on 07/23/03 under PSN-03-L35, with the problem description, effective workaround and estimated release of coding modifications to the software version impacted.
Root Cause
Other
Distribution Pattern
Nationwide
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.