Edwards Lifesciences Swan-Ganz Thermodilution Catheter with AMC THROMBOSHIELD (An Antimicrobial * Heparin Coating) REF: 139HF75P For use inpatients who require hemodynamic monitoring. Intended to be used in combination with clinical pressure monitoring equipment to measure right heart and pulmonary artery pressures, and with the Vigilance Monitor to measure cardiac output.

FDA Recall Z-1240-2009 · Posted 2009-04-17

Recall StatusTerminated
Product CodeKRA
Date Initiated2009-02-09
Date Terminated2011-04-11
Product Quantity2,728 units

Reason for Recall

Cardiac output measurements were unavailable and the following message is displayed in the Vigilance monitor: "Fault. CCO Catheter Verification. Use Bolus Mode".

Firm Action

Edwards initiated the worldwide recall via Urgent Customer Notification letters dated February 9, 2009 and are being sent by Federal Express starting on February 10, 2009. These letters are requesting that customers cease using Swan-Ganz" Continuous Cardiac Output (CCO) Thermodilution Catheter products from the indicated lots and that they return all unused product. Edwards Lifesciences requested that the customer provide written verification of stock and identify any unused product to be returned. Questions should be directed to Technical Service at 800-822-9837.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution -- AL, AR, AZ, CA, CT, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NJ, NM, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Hawaii, and Puerto Rico and countries of Canada, Austria, Belgium, Switzerland, Germany, Denmark, Spain, Finland, France, UK, Greece, Ireland, Israel, Italy, Kuwait, Netherlands, Norway, Portugal, Sweden, Colombia, Australia, and Singapore.