Open, Classified

Nester Embolization Microcoil RPN GPN MWCE-18-5-2-NESTER G52733 MWCE-18-5-3-NESTER G52734 MWCE-18-7-3-NESTER-01 G47338 MWCE-18-7-2-NESTER-01 G47337 MWCE-18-7-5-NESTER-01 G47340 MWCE-18-3-2-NESTER G52731 MWCE-18-7-8-NESTER-01 G47342 MWCE-18-7-8-NESTER G52739 MWCE-18-7-5-NESTER G23521 MWCE-18-3-3-NESTER G52732 MWCE-18-5-3-NESTER-01 G47334 MWCE-18-7-6-NESTER G52738 MWCE-18-7-6-NESTER-01 G47341 MWCE-18-3-3-NESTER-01 G47332 MWCE-18-7-4-NESTER G52737 MWCE-18-7-3-NESTER G52736 MWCE-18-3-2-NESTER-01 G47331 MWCE-18-5-2-NESTER-01 G47333 MWCE-18-2-2-NESTER-01 G47344 MWCE-18-5-5-NESTER-01 G47335 MWCE-18-7-2-NESTER G52735 MWCE-18-7-4-NESTER-01 G47339 MWCE-18-7-10-NESTER G52741 MWCE-18-5-5-NESTER G23520

FDA Recall Z-1241-2022 · Posted None

Recalling FirmCook Incorporated
Recall StatusOpen, Classified
Product CodeKRD
Date Initiated2022-05-05
Product Quantity106,033

Reason for Recall

The loading cartridges included with these devices may contain a small, unintended, stainless-steel cannula.

Firm Action

On May 5, 2022, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were advised that the presence of a small stainless steel cannula inside the loading cartridge may go undetected by the user. Customers were instructed to examine their inventory and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. Customers should return the product to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Please share this information with appropriate personnel, down to the user level, within your organization or with any organization where the affected devices may have been transferred. Physicians should practice their standard patient monitoring following the procedure for an early identification of any complications to mitigate their severity. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.

Root Cause

Under Investigation by firm

Distribution Pattern

Domestic distribution nationwide. Foreign distribution worldwide.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.